Mouse EPO/Erythropoietin ELISA Kit
Validation Data Gallery
Product Information
KE10031 is a solid phase sandwich Enzyme Linked-Immuno-Sorbent Assay (Sandwich ELISA). The Erythropoietin ELISA kit is to be used to detect and quantify protein levels of endogenous Erythropoietin. The assay recognizes mouse Erythropoietin. An antibody specific for Erythropoietin has been pre-coated onto the microwells. The Erythropoietin protein in samples is captured by the coated antibody after incubation. Following extensive washing, another antibody of biotinylated specific for mouse Erythropoietin is added to detect the captured mouse Erythropoietin protein. For signal development, Streptavidin-HRP is added, followed by Tetramethyl-benzidine (TMB) reagent. Solution containing sulfuric acid is used to stop color development and the color intensity which is proportional to the quantity of bound protein is measurable at 450 nm with the correction wavelength set at 630 nm.
| Product name | Mouse EPO/Erythropoietin ELISA Kit |
| Tests | 1 X 96 well plate |
| Sample type | Serum, Plasma, Cell culture supernatants 备注:对未验证样本,我们提供技术支持和售后保障。 |
| Assay type | Sandwich |
| Sensitivity | 12 pg/mL |
| Range | 78.1-5000 pg/mL |
| Reactivity | Mouse |
| Tested applications | Sandwich ELISA |
| Gene ID (NCBI) | 13856 |
Recovery
| Sample Type | Average | Range |
|---|---|---|
| Mouse serum | 90% | 71%-117% |
| cell culture supernatants | 98% | 70%-119% |
IntraAssay
| Sample | n | mean ( pg/mL) | SD | CV% |
|---|---|---|---|---|
| 1 | 20 | 252.2 | 20.7 | 8.2 |
| 2 | 20 | 1,053.8 | 47.4 | 4.5 |
| 3 | 20 | 4,276.6 | 197.0 | 4.6 |
InterAssay
| Sample | n | mean ( pg/mL) | SD | CV% |
|---|---|---|---|---|
| 1 | 24 | 267.7 | 22.8 | 8.5 |
| 2 | 24 | 1,074.6 | 48.0 | 4.5 |
| 3 | 24 | 4,481.0 | 204.8 | 4.6 |
Background Information
Erythropoietin (EPO) is a glycoprotein hormone that regulates the production of red blood cells and biosynthesis of hemoglobin. The predominant expression of this gene shifts from the liver during fetal development to kidney in adults, and and the secreted protein will travel through the blood stream to reach to the bone marrow to stimulate hematopoietic stem cell differentiation to RBC. EPO binds to the cognate EPO receptor (EPOR) on erythroid progenitor cells, thus preventing apoptosis and stimulating their differentiation and maturation into erythrocyte. However, EPO protein and its receptors have also been shown to be cytoprotective in extra-hematopoietic tissues including the retina tissue. A complete lack of erythropoietin causes embryonic lethal anemia in mice. The conditional inactivation of erythropoietin in adult mice results in a chronic, normocytic and normochromic anemia.
Properties
| Storage Instructions | All the reagents are stored at 2-8℃ for 6 months or -20℃ for 12 months. Refer to the protocol for further storage instructions. |
| Synonyms | Epo, erythropoietin |
FAQ
1、我的样本稀释液有限,如何处理稀释倍数较大的样本?
我们建议您对样本进行梯度稀释。例如,人血清样本需要1000000倍稀释时,您可以分3个梯度,每个梯度100倍,进行稀释:
第一步:取2 uL人血清样本,加入198 uL样本稀释液,混匀得到100倍稀释的人血清样本
第二步:取2uL 100倍稀释的人血清样本+ 198 uL样本稀释液进行稀释,混匀得到10000倍稀释的人血清样本
第三步:取2uL 10000倍稀释的人血清样本+ 198 uL样本稀释液进行稀释,混匀得到1000000倍稀释的人血清样本。
如果样本需要进一步稀释,您可以使用更多梯度。除了我们推荐的样本稀释液,不建议您使用其他溶液或者水稀释。如果您的样本稀释液仍然不足,请联系我们的客服为您提供更多样本稀释液。
Publications
| Species | Sample Type | Title |
|---|---|---|
Nutrition Changes in iron metabolism centered on hepcidin due to high-intensity exercise under restricted food intake. | ||
ACS Pharmacol Transl Sci Prolyl Hydroxylase Domain Protein Inhibitor Not Harboring a 2-Oxoglutarate Scaffold Protects against Hypoxic Stress. | ||
Lipids Health Dis Deficiency of lysophosphatidic acid receptor 3 decreases erythropoietin production in hypoxic mouse kidneys | ||
Physiol Rep Pharmacological prolyl hydroxylase inhibition enhances high-intensity exercise capacity in mice | ||
Research (Wash D C) Directed Evolution of T7 RNA Polymerase Minimizes dsRNA By-product and Enables High-Fidelity mRNA Synthesis for Demanding Therapeutic Applications. |